10. FREQUENTLY ASKED QUESTIONS
For integrated CPAP devices:
a. When I wake up in the morning, my nose and throat feel dry. What can I do?
Try increasing your humidity setting. If this does not help, please contact your clinician for advice.
b. How do I prevent condensation in the tubing?
The humidity setting enables adjustment of humidity, so that there should be fine misting in the six inches (183cm) of the
breathing tube closest to your face. If the humidity setting is too high for the conditions, condensation may occur in the
breathing tube.
Note: Using greater than 6' (183 cm) of tubing will increase condensation.
There are several ways of reducing condensation in the tube:
i Ensure the device is not positioned in a cool draft.
ii Decrease the humidity setting on your device slightly. Gradually decrease the setting until the condensation no
longer occurs.
iii Warm the air in the room.
c. Do I have to use distilled water with my device?
The use of distilled water will maximize the life of the water chamber. Water from the faucet will often contain minerals
which can damage the chamber, causing pitting in the base, even if the water is passed through a filter.
d. When do I replace my water chamber?
Replace the chamber if the plastic walls of the chamber become cracked or cloudy or the chamber base becomes pitted.
(Note: If the chamber leaks at all, it should be replaced immediately).
Can I use my device in other countries?
Yes. Simply use the appropriate electrical socket adaptor and the device will automatically adapt to any voltage supply.
Note: When moving your device, ensure that the water chamber is empty. Machine failure due to water damage is not covered by
warranty.
Can I use supplemental oxygen with my device?
Yes, oxygen can be administered at the mask. Turn the device on before turning on the oxygen. Ensure that the oxygen source is
turned off before turning off the device to avoid oxygen accumulating in the machine.
DIMENSIONS: 275mm x 170mm x 140mm
(10.8" x 6.7" x 5.5")
WEIGHT: 1.9kg (4.2lbs)
(3.0kg packaged in bag incl. accessories)
ELECTRICAL RATINGS
Supply Frequency: 50-60 Hz
Supply Voltage and Current:
100-115V~ (1.2A, 1.3A max),
220-240V~ (0.8A, 1.8A max)
Note: These values represent
average current
DC to AC Inverter requirements: 115V - 200W/300W surge
230V - 300W/500W surge
Heater Plate : 85W max
Heater Plate Temperature: 65°C (149°F) max
PERFORMANCE
Pressure Range 4 to 20
cmH
2
O
Altitude Range 0 to 3000m, or 0 to 9000 ft
STANDARDS COMPLIANCE
Designed to conform to the
requirements of: EN / IEC 60601-1
UL60601-1
CSA C22.2/No.601.1
AS3200.1.0
The device complies with the electromagnetic compatibility
requirements of IEC 60601-1-2. In certain circumstances, the
device may affect or be affected by nearby equipment due to
the effects of electromagnetic interference. If this should
happen, try moving your unit or the location of the device
causing interference, or alternatively consult your homecare
provider.